Dial gauge N male (50 /75 Ω) product photo Front View L
For a larger view click on the picture.

Dial gauge N male (50 /75 Ω)

537011
Compare

RF characteristics

Number of interfaces:
1
Interface type:
N male (50 Ω) per IEC 61169-16
N male (75 Ω) per IEC 61169-16

Mechanical characteristics

Accuracy level 1:
Grade 0
Accuracy level 1 tolerance range:
0.08 mm
Accuracy level 1 pin offset:
5.28 to 5.36 mm
Accuracy level 2:
Grade 2
Accuracy level 2 tolerance range:
0.50 mm
Accuracy level 2 pin offset:
5.33 to 5.83 mm
Gauge range:
5 mm
Scale interval:
0.01 mm
Measurement accuracy:
0.005 mm
Material:
stainless steel (dial gauge), stainless steel (reference gauge)
Surface Coating:
no finish (reference gauge), no finish (dial gauge)
Degree of protection:
IP 40
Weight, approx.:
300.0 g

Environmental conditions

Operating temperature:
+18 to +28 °C
Relative humidity, max.:
95% (condensation not allowed)
Storage ambient temperature:
-30 to +50 °C
Storage relative humidity, max.:
95% (condensation not allowed)

Scope of delivery & accessories

Delivery scope:
dial gauge, reference gauge, storage box, handling instructions, factory calibration certificate

Further remarks

Customs tariff number:
90318020

Downloads

537011-DS.pdf

Product data sheet – 537011 – 537011-DS.pdf

M31004.pdf

Handling instructions – 537011 – M31004.pdf